Subject: uVax3 back in acq
Nicola.Zaccheo@lngs.infn.it
Date: Fri Mar 10 2000 - 07:19:12 EST
Ciao a tutti,
during run 19218 a (probably) electric-power glitch caused two
problems:
- the LeCroy HV system that lost power on some channels
of SM5 (immediatly fixed after a power cycle of the crate number
7 where is the CAMAC interface module);
- a TOHM lower LAMBDA power supply on SM5 broke. This power supply
showed -8.0 V on -12V line (with a 50mV AC component). For this reason
we excluded uVax3 from acquisition.
Today we replaced this power supply using a different one (we have no
LAMBDA spares available!). After a 40 min. mini-acq we put back uVax3 in
acquisition (starting from run 19226).
A presto,
Nicola