Nicola.Zaccheo@lngs.infn.it
Date: Fri May 12 2000 - 06:58:11 EDT
Ciao a tutti,
during run 19527 (still running) I noticed a lot of no Q's responses
from M:3,B:2,C:5,N:22 - InterERP TDC on uVax3.
Looking the start and the clear signals with the digital scope,
we found that there was no Start signal when no Q's were generated...
...this suggested us that something wrong was in the logic that
provides the OR between the Common Start and the local Ehigh signals.
In fact the problems disappeared unplugging and plugging back the
Ehigh3 input on the InterERP Timing Manager Module.
Last no Q received at 11:32AM UTC time.
A presto,
Nicola